DE-5004 LCRメータ
True RMS
Resistance; Frequency; Continuity; Diode; LCR (Inductance Capacitance Resistance)
Intergrate multimeter & L.C.R. function in one meter
Dual display (6000 counts -Main ; 2000 counts-sub. )
-
Measurement Functions:ACV, DCV, ACA, DCA, DCμA, ACμA, DCmA, ACmA, Resistance; Frequency; Continuity; Diode; LCR (Inductance Capacitance Resistance)
-
LCD Display:6600 Counts (main)
2000 Counts (sub) -
Sampling Rate:3 Times/Sec. (DMM mode)
1 Time/Sec. (LCR mode) - Range:Automatic & Manual
- Overload:”OL” Display On LCD
-
Low Battery:
Display On LCD
-
LCR Measurement (auto range):
*Rs, Rp, Cs, Cp, Ls, Lp, phase angle, Dissipation factor, Quality factor
*Auto select series or parallel test mode
*Auto select 100Hz/1kHz/10kHz test frequency, 0.5V rms approx.
*Calibration - Auto Power Off:In 15 Minutes Approx.
- Battery:AA 1.5V battery x 2
- Safety standard:IEC 61010-1 CAT III 600V , IEC 61010-1 CAT.II 1000V
Measurement Ranges:
23°C ± 5°C max. 80%R.H. Accuracy: ±(%rdg +dgt)
Measurement |
Range |
Resolution |
Accuracy (Responding Hz ) |
Input Impedance |
Max. Input Voltage |
|
DE-5003(sine wave) |
DE-5004*(True RMS) |
|||||
ACV |
6.000V |
0.001V |
0.7%+6(40Hz~1kHz) |
0.7%+6(40Hz~1kHz) |
Approx. 11MΩ |
1000V rms |
60.00V |
0.01V |
0.7%+5(40Hz~1kHz) |
0.7%+6(40Hz~1kHz) 2.0%+6(1kHz~5kHz) |
Approx. 10MΩ |
||
600.0V |
0.1V |
|||||
1000V |
1V |
1.2%+5(40Hz~1kHz) |
1.2%+6(40Hz~1kHz) 2.0%+6(1kHz~2kHz) |
*True RMS: valid from 10%~100% of range
Measurement |
Range |
Resolution |
Accuracy |
Input Impedance |
Max. Input Voltage |
|
DE-5003 |
DE-5004 |
|||||
DCV |
6.000V |
0.001V |
0.1%+3 |
Approx. 11MΩ |
1000V rms |
|
60.00V |
0.01V |
0.2%+3 |
Approx. 10MΩ |
|||
600.0V |
0.1V |
0.4%+3 |
||||
1000V |
1V |
0.6%+5 |
||||
|
|
|
|
|
|
|
VA+Hz ( at secondary display) |
Accuracy |
Max. Input Voltage |
||
Range |
Resolution |
DE 5003 |
DE 5004 |
|
600.0Hz |
0.1Hz |
0.1%+5 |
1000V rms |
|
6.000KHz |
0.001KHz |
|||
60.00KHz |
0.01KHz |
|||
|
|
|
|
|
Measurement ADP |
Range |
Resolution |
Accuracy (Responding Hz ) |
Input Impedance |
Max. Input Voltage |
|
DE-5003(sine wave) |
DE-5004*(True RMS) |
|||||
ACmV |
600.0mV |
0.1mV |
0.7%+5(40Hz~2KHz) |
0.7%+5(40Hz~10KHz) |
Approx. 10MΩ |
1000V rms |
DCmV |
600.0mV |
0.1mV |
0.2%+5 |
0.2%+5 |
*True RMS: valid from 10%~100% of range
Measurement |
Range |
Resolution |
Accuracy |
Max. Input Voltage |
Ω Resistance |
600.0 Ω * |
0.1Ω |
0.5%+5 |
1000V rms |
6.000KΩ |
0.001KΩ |
|||
60.00KΩ |
0.01KΩ |
|||
600.0KΩ |
0.1KΩ |
|||
6.000MΩ |
0.001MΩ |
1.0%+5 |
||
60.00MΩ |
0.01MΩ |
2.5%+5 |
Measurement |
Range |
Resolution |
Remarks |
Open-loop Voltage |
Max. Input Voltage |
Continuity |
600.0Ω |
0.1Ω |
Buzzer emits at around ≦10Ω±3Ω |
Approx. -3.3V |
1000V rms |
Measurement |
Range |
Resolution |
Accuracy |
Open-loop Voltage |
Max. Input Voltage |
Diode |
6.000V |
0.001V |
0.5%+5 |
Approx. 2.8V |
1000V rms |
Measurement |
Range |
Resolution |
Accuracy |
Remarks |
Max. Input Voltage |
Hz |
600.0Hz |
0.1Hz |
0.1%+5 |
Available reading of Duty: 5.0%~94.9%(<600Hz) 10.0%~90.0%(<6kHz) 20.0%~80.0%(<60kHz)
Resolution 0.1% The UL will be show on LCD. Duty larger than 94.9%, The OL will be show on LCD. |
1000V rms |
6.000kHz |
0.001kHz |
||||
60.00kHz |
0.01kHz |
||||
600.0kHz |
0.1kHz |
||||
6.000MHz |
0.001 MHz |
||||
20.00MHz |
0.01MHz |
Measurement |
Range |
Resolution |
Accuracy |
Overload Protection |
DC μA |
600.0μA |
0.1μA |
0.5%+5 |
630mA/1000V fuse |
6000μA |
1μA |
|||
DC mA |
60.00mA |
0.01mA |
||
600.0mA |
0.1mA |
|||
DC A |
*10.00A |
0.01A |
1.0%+10 |
10A/1000V fuse |
*overload 11 ~ 20A stands with 30 sec. max.
Measurement |
Range |
Resolution |
Accuracy ( Responding Hz) |
Overload Protection |
|
DE-5003(sine wave) |
DE-5004**(True RMS) |
||||
AC μA |
600.0μA |
0.1μA |
1.0%+10(40Hz~1kHz) 1.5%+10(1KHz~2KHz) |
1.0%+10(40Hz~1kHz) 1.5%+10(1KHz~5KHz) |
630mA/1000V fuse |
6000μA |
1μA |
||||
AC mA |
60.00mA |
0.01mA |
|||
600.0mA |
0.1mA |
||||
AC A* |
10.00A |
0.01A |
1.5%+10(40Hz~2kHz) |
1.5%+10(40Hz~5kHz) |
10A/1000V fuse |
*over load 11 ~ 20A stands with 30 sec. max.
**True RMS: valid from 10%~100% of range
LCR Inductance
Ls / Lp |
0.3∼60.0μH |
60.1∼600.0μH |
601∼6000μH |
6.01∼60.00mH |
60.1∼600.0mH |
601∼6000mH |
6.01∼60.00H |
60.1∼100.0H |
Resolution |
0.1μH |
0.1μH |
1μH |
0.01mH |
0.1mH |
1mH |
0.01H |
0.1H |
Accuracy |
±(3.0%+20) |
±(1.5%+5) |
±(1.0%+5) |
±(1.0% +5) |
±(1.0%+5) |
±(1.0%+5) |
±(1.5%+5) |
±(3.0%+10) |
LCR Capacitance
Cs / Cp |
3.0∼600.0pF |
601∼6000pF |
6.01∼60.00nF |
60.1∼600.0nF |
601∼6000nF |
6.01∼60.00μF |
60.1∼600.0μF |
0.60∼6.00mF |
Resolution |
0.1pF |
1pF |
0.01nF |
0.1nF |
1nF |
0.01μF |
1μF |
0.01mF |
Accuracy |
±(2.0%+20) |
±(1.0%+5) |
±(0.3%+5) |
±(0.3%+5) |
±(0.5%+5) |
±(0.8%+5) |
±(1.5%+10) |
±(3.0%+20) |
LCR Resistance
Rs / Rs |
0.01∼60.00Ω |
60.1∼600.0Ω |
601∼6000Ω |
6.01∼60.00KΩ |
60.1∼600.0KΩ |
601∼6000KΩ |
6.01∼20.00MΩ |
Resolution |
0.01Ω |
0.1Ω |
1Ω |
0.01KΩ |
0.1KΩ |
1KΩ |
0.01MΩ |
Accuracy |
±(1.0%+20) |
±(0.3%+5) |
±(0.3%+5) |
±(0.3%+5) |
±(0.3%+5) |
±(1.0%+5) |
±(3.0%+5) |
*Above accuracy is guaranteed after open/short calibration.
*Some types of inductor, the inductance value changes acutely according to the test voltage applied.
*The meter provides the output voltage around 2.2kΩ / 0.6Vrms at sine wave. However, the test voltage is altered in according to the resistance of the inductor be tested.
*Temperature coefficient:0.1 x (specified accuracy) / ℃ (from 0℃ to 18℃ or 28℃ to 50℃)
TEMP./Humidity:
- Operation:0°C~ 50°C(32℉~122℉)/<80% R.H.(no condensation)
- Storage:-10°C~ 60°C(14℉~140℉)/<70%R.H.(no condensation)
Dimension & Weight:
- 181(L)x92(W)x57(H)mm
- Approx.400g (excluding battery)
Accessories:
- 1.5V AA Battery x 2
- Instruction Manual x 1
- Test Lead (black + red) x 1
-
True RMS
- Integrate Multimeter & L.C.R Function In One Meter
- Data Hold
- Dual Display
- Max./Min Hold
- Backlight
- Auto Range & Manual Range
- Auto Power Off Selection